Menü
Materialien Maschinen
Einblicke
Artikel Videos Problem & Lösung
About
Über Chembull Medienpartner
Kontaktieren Sie uns
Sprache
EnglishالعربيةPortuguêsEspañolDeutschTürkçe한국어
Zurück zu Problem & Lösung
Material Drying 6 Aufrufe

PA66 medical device boundaries

Problembeschreibung
PA66 medical parts have sharp strength drop and dimension out of control after repeated steam sterilization
Ursachenanalyse
PA66's moisture absorption characteristic makes it absorb water, swell, and hydrolyze during high-temperature steam sterilization, and performance cumulatively deteriorates with repeated sterilization; medical scenarios have higher requirements for batch consistency and traceability
Lösung
①For repeated steam sterilization scenarios, switch to hydrolysis-resistant or other materials; ②Verify disposable products by single sterilization; ③Batch traceability of medical-grade grades; after adjustment, take continuous samples to verify the effect, confirm stability before resuming mass production
Wichtigste Erkenntnis
PA66's moisture absorption nature limits its medical sterilization applications; grade hydrolysis resistance level and usage count limitation are the feasible boundaries; after confirming effectiveness, fix the parameters and practices into the work instructions to prevent recurrence
PA66;medical;steam sterilization;hydrolysis;traceability

War das hilfreich?

0 Personen fanden dies hilfreich

ChemAI